VOC Testing Chamber Model No.: VWH-60

VOC Testing Chamber  Model No.: VWH-60

Model No.︰VWH-60型

Brand Name︰-

Country of Origin︰China

Unit Price︰-

Minimum Order︰1 pc

Inquire Now

Product Description

1.Application
This product is used to testing the VOC emissions of wood-based panel, water based interior wall coating materials,textile etc. light industry products.

 

2.Standards     

T/310101002-C003-2016

 <Limit of harmful substances of plastic and rubber sports surface for school >

JG/T 481-2015    

<Low Volatile Organic Compounds (VOC) Water Based Interior Wall Coating Materials>

ASTM D6007—02

<Standard Test Method for Determining Formaldehyde Concentrations in Air from Wood Products Using a Small-Scale Chamber1>

JG/T****

<Chamber test method for determining VOC emission rates from building decorating materials>(not yet released)

 

3.Technical Parameters

No.

Items

Technical Parameters

 

1

Temp

Range:15-65,Error Tolerance:±0.5℃

2

Humidity

Range:5-80%,Error Tolerance:±1%

3

Exchange Rate

(0.2-2)pcs/h, Error Tolerance:±1%

(Scope of exchange adjustment:12L-120L/h)

4

Chamber Volume

60±0.02LRemarks: Size will be made as request

5

Air flow rate

(0.1—0.3)ms   Continuous Adjustment

Error Tolerance:≤0.05m/s

6

Internal pressure

10±5Pa

7

Tightness

1KPa,Static Leak Rate≤1×10-3 m3/min

8

Recovery rate

80%,(Toluene,C12H26

 

9

Construction

(1.)Test chamber in all three walls:

inner wall-mirror stainless steel

middle-rock wood insulation wall

outer wall-spray material

  1. )Non-releasing,non-adsorbing materials in the chamber

(3.)There has air-inlet,testing,sampling at chamber. Air-inlet connect with regulation valve of common clean air source. 

 

 

 

 

 

 

 

 

10

Common clean air source

Common air source is composed of turbo fan motor,particle filter, VOC filter, formaldehyde filter,humidity generator,constant temp control unit, air flow control unit and so on.

Air flowtotal2 m³/h

Cleanliness:

 concentration of formaldehyde≤0.005mg/m³

Single concentration of VOC≤0.002mg/ m³

Concentration of TVOC≤0.02mg/ m³

Dew-point temp5℃--40℃

The air distributor divided into multiple channels, be controlled respectively.

11

Control system

Display:temp,humidity,time,air flow,pressure

12

Key components

Import brand:compressor,temp &humidity sensor,temp controller and flow-meter...